Monday, January 28, 2013

New filament continued Jan 2013

Monday

Gun on Monday morning after mysterious shutdown over the weekend.

Initial extractor current 84.8 uA, was 108 uA with beam off Friday afternoon.

1 kV, 73.8 uA, gun vac jumps to 1.8e-9 and rising slowly, 156 pA for 30 um standard aperture

After 30 min Iext up to 99.6, gun 4e-10
After 70 min, 88 uA at 5 kV, 2.25e-9 vac

Noon, ext 108.5 beam off, 7e-10 vac

Around 2:30 gun shuts down. This time got error "519 EHT Set Extractor Trip"

At 3:15 gun shut down again, no clear reason, no error message.

Gun runs through the afternoon. EHT off now 114.2, was around 94.5 @ 1 kV.

Turn on EHT 1 kV, see if it runs through the night. Now 8:30 pm.

Tuesday

This time gun did not shut down. Ext current relatively stable, starts 94.7 ends 95.5. Gun vac drops slightly, starts 3.0e-9 ends 2.7e-9

Gun off 115.3 uA, pressure 1.0e-9 within a minute.

1 comment:

  1. Hello!
    We have a similar problem with the awesome gun shutdown.
    How did it end in your case?

    ReplyDelete